Shorter crowns showed more localized implant stress in a model

A peer-reviewed computer study found the 8 mm configuration had the highest modeled stress peaks, but the affected areas were smaller.

A shorter crown configuration showed the most concentrated stress pattern in a computer model of a dental implant, although it also had the highest peak stress values in the implant and its zirconia support. The finding came from a finite-element analysis of modeled structures. It does not show that shorter crowns improve implant survival or prevent complications in patients.

The study compared three simulated versions of a lower-left premolar rehabilitation. Each used an external-hexagon titanium implant measuring 4 by 11 mm, a zirconia abutment and a lithium disilicate crown. The models differed in crown height and the resulting crown-to-implant ratio.

What the models showed

The reduced-size, or RS, model used an 8 mm crown and a crown-to-implant ratio of 0.72:1. The standard-size, or SS, model used 10 mm and 0.90:1, while the large-size, or LS, model used 12 mm and 1.1:1. The three configurations were then compared.

For the zirconia abutment, the RS model reached a maximum von Mises stress of 188.63 MPa, compared with 177.87 MPa in both SS and LS. Von Mises stress was the measure used for the abutment and implant comparisons. The RS stress was concentrated over a smaller area.

The implant showed a similar pattern. Its modeled maximum stress was 336.23 MPa in RS, versus 324.55 MPa in both SS and LS. In the RS model, the stress occupied a smaller area near the first implant thread, while the SS and LS patterns covered larger areas.

The reported bone values also differed between the reduced model and the other two. In cortical bone, the RS field had reported extrema of -8.69 and -82.78 MPa, compared with -10.59 and -97.06 MPa for SS and LS. In trabecular bone, the corresponding RS values were 1.99 and 0.44 MPa, compared with 2.35 and 0.47 MPa for SS and LS. The bone analysis used maximum and minimum principal stresses, the model's directional stress values.

The standard and large configurations produced the same reported values in the main abutment and implant comparisons. The study's abstract also said no differences in stress-distribution values were observed between SS and LS.

A controlled computer experiment

The researchers applied a 200 N computational biting load straight along the axis of the modeled structures. They treated cortical and trabecular bone as isotropic, homogeneous and linearly elastic, with complete osseointegration between implant and bone. In plain terms, the simulation gave the materials uniform properties and assumed the implant was fully integrated with the surrounding bone.

The study used models rather than measurements from people or animals. No inferential statistical analysis was performed. The authors considered the converged finite-element results stable and reproducible, so the comparison rests on the model outputs rather than estimates of variation across independent samples.

What the result does not establish

The authors rejected their null hypothesis, which proposed that the crown-to-implant ratio would not affect the system's biomechanical performance. They judged the 8 mm RS configuration to have the better modeled distribution of stress and resistance to tensile and compressive forces. That interpretation is limited to the simulation and does not by itself establish clinical benefit.

The study's clinical reach is limited by its simplified setup. It used a linear simulation with homogeneous, linearly elastic bone, applied only axial loading and did not include in vitro step-stress testing to check how the models behave under repeated force. Stress peaks at the implant-abutment and implant-bone interfaces were also removed from the output, meaning the reported maxima do not represent the untrimmed interface peaks.

The simplified setup means the results should not be treated as a clinical prediction. The paper did not include in vitro step-stress validation, so it did not test the modeled pattern outside the computer simulation. The authors' interpretation is therefore limited to the reported model.

The datasets generated or analyzed for the study are available from the corresponding author on reasonable request. The authors reported no commercial or associative conflict of interest connected with the manuscript.

Paper data and sources

Original title: Zirconia abutments and reduced crown-to-implant ratios: Insights into stress distribution
Authors: GRASSI L, SILVA A, BORGES G et al.
Journal/Repository: Brazilian oral research
Status: Peer-reviewed
First online: 2026-10-19
DOI: Not available
Original paper · Full text